Guralp-systems CMG-3ESPCD Uživatelský manuál Strana 37

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 98
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 36
CMG-3ESPCD Digital Broadband Seismometer Calibrating the 3ESPCD
The bbnoisecal script automatically performs appropriate averaging
to reduce the effects of aliasing and cultural noise.
4.3.1 Sensor response codes
Sensor Sensor type code Units
(V/A)
CMG-3T or 3ESP, 30 s – 50 Hz response
CMG-3_30S_50HZ
V
CMG-3T or 3ESP, 60 s – 50 Hz response
CMG-3_60S_50HZ
V
CMG-3T or 3ESP, 100 s – 50 Hz response
CMG-3_100S_50HZ
V
CMG-3T or 3ESP, 120 s – 50 Hz response
CMG-3_120S_50HZ
V
CMG-3T or 3ESP, 360 s – 50 Hz response
CMG-3_360S_50HZ
V
CMG-3TB/3V/3ESP borehole, 30 s – 50 Hz
CMG-3B_30S_50HZ
V
CMG-3TB/3V/3ESP borehole, 100 s – 50 Hz
CMG3B_100S_50HZ
V
CMG-3TB/3V/3ESP borehole, 120 s – 50 Hz
CMG3B_120S_50HZ
V
4.4 The coil constant
The feedback coil constant K is measured at the time of manufacture, and
printed on the calibration sheet. Using this value will give good results at the
time of installation. However, it may change over time. The coil constant can
be determined by tilting the instrument and measuring its response to gravity.
To do this, you will need apparatus for measuring tilt angles accurately.
1. Measure the acceleration due to gravity, g, at your location.
2. Tilt the instrument slightly, and measure its attitude and the gain of the
mass position output for the component you wish to calibrate.
MAN-C3E-0002 37 Issue B - November 2013
Zobrazit stránku 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 97 98

Komentáře k této Příručce

Žádné komentáře